Stroke is a widespread and perilous condition globally, necessitating effective
rehabilitation assessment methods. Presently, clinicians rely on manual observation or
costly equipment like motion capture systems to evaluate stroke patients’ recovery
progress. This study introduces a streamlined measurement system designed for assessing
lower limb function recovery in stroke patients. The system employs wearable IMU
(Inertial Measurement Unit) sensor modules strategically placed on the lower limbs to
capture joint angles. An algorithm processes sensor data to calculate precise joint angles.
To counteract environmental interference affecting IMU sensors, a Kalman filter is
implemented to minimize errors, providing real-time adjustments for sensor accuracy.
Experimental validation, using exercises and criteria from the FUGL-MEYER Assessment,
a widely adopted method for evaluating stroke patients’ physical performance, was
conducted to compare the proposed system with the VICON motion capture system. Results
demonstrated close alignment between the two systems, affirming the proposed system’s
reliability, applicability, and practicality for real-life medical evaluations. Importantly, the
proposed system offers a cost-effective and portable alternative for measuring joint angles,
maintaining effectiveness and precision. In summary, this system offers a straightforward
design, easy installation, and affordable hardware, making it a viable and practical
solution for evaluating lower limb rehabilitation in stroke patients.
- Cím és alcím
- The Development of a Motion-Tracking System to Assess the Recovery Level for Stroke Survivors
- Szerző
- Nguyen, Thanh-Trung
- Nguyen, Quang-Duy
- Ha, Minh-Quang
- Hoang, Xuan-Tung
- Megjelenés ideje
- 2024
- Hozzáférés szintje
- Open access
- ISSN, e-ISSN
- 1785-8860
- Nyelv
- en
- Terjedelem
- 20 p.
- Tárgyszó
- imu sensors, kalman filter, fugl-meyer assessment, lowerlimb, vicon motion capture system
- Változat
- Kiadói változat
- Egyéb azonosítók
- DOI: 10.12700/APH.21.9.2024.9.2
- A cikket/könyvrészletet tartalmazó dokumentum címe
- Acta Polytechnica Hungarica
- A forrás folyóirat éve
- 2024
- A forrás folyóirat évfolyama
- 21. évf.
- A forrás folyóirat száma
- 9. sz.
- Műfaj
- Tudományos cikk
- Tudományterület
- Műszaki tudományok - multidiszciplináris műszaki tudományok
- Egyetem
- Óbudai Egyetem